Fluke Universal Remote Fluke 16 User Manual

®
16  
Multimeter with Temperature  
Calibration Information  
W
Warning  
To avoid electrical shock, remove test leads and any input signals from the Meter before opening the case.  
Caution  
This Meter contains parts that can be damaged by static discharge. Follow the standard practices for handling  
static sensitive devices.  
To contact Fluke, call one of the following telephone  
numbers:  
Introduction  
The 16 Calibration Information instruction sheet provides the  
information necessary to calibrate and verify the Fluke  
Model 16 Multimeter with Temperature (hereafter known as  
the Meter). This instruction sheet provides the following  
information:  
USA and Canada: 1-888-99-FLUKE (1-888-993-5853)  
Europe: +31-402-678-200  
Japan: +81-3-3434-0181  
Singapore: +65-738-5655  
Product specifications  
Anywhere in the world: +1-425-356-5500  
Disassembly and reassembly  
Performance test procedures  
Calibration procedures and adjustments  
Replacement parts  
Specifications  
Specifications are in Table 1. Accuracy is specified for a  
period of one year after calibration, at 18ºC to 28ºC (64ºF to  
82ºF) with relative humidity to 90%. AC conversions are ac-  
coupled, average responding, and calibrated to the rms  
value of a sine wave input.  
See the instruction sheet for operating instructions.  
Service  
The Meter is warranted to be free from defects in material  
and workmanship for one year. The complete warranty  
statement is in the Meter’s instruction sheet.  
Table 1. Specifications  
Maximum Voltage Between any Terminal and Earth Ground (excludes 10% tolerance) 600 V ac or dc  
Display  
3-3/4-digits, 4000 counts, updates 4/sec  
-10°C to 50°C  
Operating Temperature  
Storage Temperature  
Temperature Coefficient  
Relative Humidity  
-30°C to 60°C indefinitely (to -40°C for 100 hrs)  
0.1 x (specified accuracy)/°C (<18°C or >28°C)  
0% to 90% (-10°C to 35°C)  
0% to 70% (35°C to 50°C)  
Altitude Operation/Non-operation  
Battery Type  
2000 m  
9V, NEDA 1604 or IEC 6F22  
Battery Life  
650 continuous hours with alkaline  
450 continuous hours with carbon-zinc  
3 meter drops: Sinusoidal vibration up to 3 G @ 55 Hz  
1.35 in x 2.75 in x 5.55 in  
Shock, Vibration  
Size (HxWxL)  
(3.46 cm x 7.05 cm x 14.23 cm)  
10 oz (286g)  
Weight  
EMI Regulations  
Safety  
Complies with FCC Part 15B, Class B, EN50081-1, EN50082-1.  
Designed to Protection Class II requirement of UL1244, ANSI/ISA-S82.01 - 1988, CSA C22.2 No 231,  
and VDE 0411, and IEC 1010-1 overvoltage category III (CAT III), 600 V  
Certification  
P $  
PN 644226 July, 1997 Rev.1, 3/99  
© 1997,1999 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies.  
1
 
Specifications  
Table 1. Specifications (cont)  
Function  
Overload  
Input Impedance  
(Nominal)  
Common Mode  
Normal Mode  
Protection1  
Rejection Ratio  
Rejection2  
(1 kUnbalanced)2  
>5 M<100 pF2  
Automatic Selection  
and LoZ = >2 kΩ  
<200 pF  
600 V dc  
>60 dB at dc 50 Hz or  
60 Hz  
f
(ac coupled)3  
>10 M<100 pF2  
Automatic Selection  
and LoZ = >2 kΩ  
<200 pF3  
600 V dc  
>100 dB at dc, 50 Hz or  
60 Hz  
>50 dB at 50 Hz or 60  
Hz  
E
Open Circuit Test  
Voltage  
Full Scale Voltage  
Short Circuit  
Current  
To 4.0 MΩ  
40 MΩ  
600 V dc  
600 V dc  
<1.5 V dc  
<450 mV dc  
<1.5 V dc  
<500 µA  
J
L2  
2.4-3.0 V dc  
2.400 V dc  
0.95 mA (typical)  
Burden Voltage  
Common Mode Rejection  
Ratio  
Normal Mode  
Rejection  
600 V rms or dc  
600 V rms or dc  
<5 mV/µA  
<5 mV/µA  
>100 dB @ dc 50 Hz or 60  
Hz  
>50 dB @ 50 Hz or  
60 Hz  
µ[  
µ\  
>60 dB at dc, 50 Hz or  
60 Hz  
.
1. 3 x 106 V Hz maximum.  
2. ~2 kinput impedance up to 50 V. Impedance increases with input voltage to >300 kat 600 V.  
MIN MAX Recording Accuracy and Response Time  
Specified accuracy of the measurement function ±12 digits in dc for changes >200 ms in duration  
(±40 digits in ac). Typical 100 ms response to 80%.  
Example 1: this would mean ±1.2° when recording temperature.  
Example 2: This would mean ±12 µA when recording µA or ±12 A if used with a dc amp probe.  
MIN MAX Recording with Elapsed Time  
Elapsed Time: 0 to 100 hours (99.59)  
Resolution: 1 minute  
Accuracy: 0.3% typical  
Continuity Capture  
Detects opens or shorts of 250 µs or longer.  
3
 
16  
Multimeter with Temperature  
Disassembly and Reassembly  
Referring to Figure 1, disassemble the Meter as follows. A  
Phillips-head screwdriver and small flat-blade screwdriver  
are required.  
Cleaning  
W Warning  
To avoid electrical shock, remove test leads  
and any input signals before cleaning.  
1. Remove the test leads and set the rotary knob to OFF.  
2. Remove the Phillips-head screws from the case bottom.  
3. Separate the case top from the case bottom.  
To clean the case, wipe it with a cloth lightly dampened with  
water and a mild detergent. Do not use abrasives, solvents,  
or alcohol.  
4. To replace the battery: lift the battery from the case  
bottom and insert a new 9V battery. Be sure the positive  
and negative battery posts are oriented correctly.  
Recommended Test Equipment  
A list of recommended equipment for the performance test  
and calibration adjustment procedure is shown in Table 2.  
5. To remove the pca: insert a small, flat-blade  
screwdriver between the edge of the case top and the  
pca where shown in Figure 2. Gently unsnap a side of  
the case top from the pca. Repeat on the other side of  
the pca. Unsnap the case from the top of the pca last.  
Table 2. Recommended Equipment  
Equipment  
Minimum  
Specification  
Recommended  
Model  
AC/DC Calibrator DC Voltage: 0 to  
600 V  
5700A, or 5500A  
Lift the pca from the case top by its edges. If the  
elastomeric contact strips for the switch assembly and  
LCD are stuck to the pca, remove them without touching  
the conductive edges.  
Accuracy: ±0.25%  
AC Voltage: 0 to  
600 V  
Accuracy: ±0.5%  
Frequency: 50 to  
400 Hz  
Note  
Before reinserting the pca, make sure that the  
rotary knob is in the OFF position.  
Decade Resistor  
Resistance: 1.0 to  
40 MΩ  
Accuracy: ±0.25%  
Fluke 5500A  
Fluke 5500A  
7. To reinsert the pca: Place the pca over the four screw  
posts in the case top, then press gently on the center of  
the pca while using the small flat-edge screwdriver to  
shoehorn the pca under the snap on a side of the case  
top. Repeat on the other side and the top.  
Decade  
Capacitor  
Capacitance: 0 to  
1.000 µF  
Accuracy: ±0.5%  
8. To remove the switch support, use a small, flat-blade  
screwdriver to gently unsnap the sides and top of the  
switch support from the snaps shown in Figure 2.  
Thermocouple  
Wire K-Type  
---  
Fluke 80 PK-1  
Temperature  
Probe  
Accuracy: Certified  
to ±0.2°C ambient  
Fluke 80T-150U  
9. The LCD, switch assembly, and elastomeric contact  
strips for the LCD and switch assembly are accessible  
and can be replaced as needed. Do not allow the LCD  
to get wet. Before installing a new LCD, make sure that  
all connector contact points are clean.  
0.02°C resolution  
Princo Model ASTM  
56C  
Princo Model ASTM  
56F  
Mercury  
Thermometer  
0.05°F resolution  
Flask with cap  
---  
Dewar Flask  
Caution  
Do not touch the conductive edges of the  
elastomeric strips or the contacts on the switch  
assembly. If they are contaminated, clean them  
with isopropyl alcohol.  
9. Reassembling the Meter is the reverse of disassembling  
it. After the Meter is reassembled, execute the  
Performance Test to confirm that the Meter is working  
properly.  
4
 
Recommended Test Equipment  
Decal  
Case, Top  
LCD  
Conn.  
Elastomeric  
Conn.  
Elastomeric  
LCD to PCA  
Switch Assembly  
Actuator, Switch  
Support, Switch Assembly  
Battery  
Shock Absorber  
Case, Bottom  
Main PCA  
Shield, Bottom  
Screw, THD Form, PH.P.STL,  
5-14 x .750  
Foot  
ig3f.eps  
Figure 1. Model 16 Disassembled Unit  
5
 
16  
Multimeter with Temperature  
3
2
Snap  
Snaps  
1
Snap  
ig2f.eps  
Figure 2. Removing and Reinserting the Printed Circuit Assembly  
6
 
Performance Test  
7. Set the 5700A using the input voltage shown in Table  
4. The reading on the display should be within the  
values shown in Table 4.  
Performance Test  
W
Warning  
To avoid electric shock, do not execute the  
performance tests procedures unless the  
Meter is fully assembled.  
8. Repeat step 3 to verify that the ice bath temperature is  
at 0.0°C ±0.3°C.  
To ensure that the Meter is working properly and performs  
to specifications, use the following procedures. If the Meter  
fails this test, it needs calibration adjustment or repair.  
Calibration Adjustments  
Perform the following calibration adjustment procedures if  
the Meter fails the performance test.  
Multimeter  
Initial Steps  
1. Remove any input signals to the Meter.  
Note  
The following performance test is for all functions  
except for temperature. Temperature performance  
test follows this procedure.  
2. Remove the four screws on the back and separate the  
case bottom and case top.  
Notice that when you do so the battery remains in the  
case top and power to the Meter is disconnected.  
1. Referring to Table 3, put the Meter in the function and  
range shown for Test 1.  
Note  
2. Apply the input from the appropriate source to the [+]  
and COM jacks on the Meter. The reading on the  
display should be within the MINIMUM and MAXIMUM  
values shown in Table 3.  
To avoid stretching or bending the battery  
contacts, connect leads to the base of the  
contacts as shown in Figure 4.  
3. Test the remaining functions and ranges. Repeat steps  
1 and 2 for test numbers 2 through 41.  
3. Observing correct polarity, connect a 9 V battery to the  
battery contacts using easy hook jumpers or alligator  
clip leads (see Figure 4).  
Temperature  
Multimeter  
Note  
Note  
The performance test for the other functions must  
be done before this test is done.  
The following calibration adjustment  
procedure is for all functions except for  
temperature. Temperature calibration  
adjustment follows this procedure.  
1. Connect the equipment as shown in Figure 3.  
2. Insert a mercury thermometer to the same depth as the  
thermocouple wire to verify that the ice bath  
temperature is 0.0°C.  
1. Set the calibrator for 0 V dc. Put the Meter in the 4.000  
V dc range.  
3. Short (or apply 0.0 mV) to the copper wires at the  
5700A Calibrator.  
2. Connect the calibrator to the [+] and COM jacks on the  
Meter.  
4. Verify that the Meter reads 0.0°C ±0.8°C.  
3. Apply an input of +4.000 V dc ±0.25%  
5. If this test fails, reposition the thermocouple wires and  
mercury thermometer in the ice bath and repeat steps  
3 and 4. If the Meter continues to fail, use steps 6  
through 8 to verify the temperature performance tests.  
The Meter display should read between 3.997-4.003 V.  
If it does not, adjust R4 as described in step 4 (see  
Figure 4).  
4. Adjust R4 (see Figure 4) so that the Meter display  
reads between 3.997 and 4.003 V.  
6. Remove the short applied in step 3.  
5. Proceed to adjust the Temperature function.  
7
 
16  
Multimeter with Temperature  
Table 3. Performance Tests  
Switch  
Meter Range  
Display  
Display  
Test No.  
Input To Meter  
Position  
Minimum  
Maximum  
1
2
3
4
5
4.000 V ac  
4.000 V ac, MIN MAX  
4.000 V ac  
4.000 V ac  
4.000 V ac  
4.000 V ac  
40.00 V ac  
40.00 V ac  
400.0 V ac  
600 V ac  
4.000 V dc, MIN MAX  
4.000 V dc  
40.00 V dc  
400.0 V dc  
4.000 V dc  
40.00 V dc  
400.0 V dc  
600 V dc  
0 V  
0 V  
Center  
Center  
Center  
Center  
Right  
Center  
Center  
Right  
0.000  
0.000  
0.007  
0.007  
3.430  
3.430  
34.30  
34.30  
343.0  
586  
-0.012  
-0.002  
00.01  
-000.3  
+3.466  
-35.33  
+99.0  
-606  
0.003  
0.040  
0.013  
0.013  
3.570  
3.570  
35.70  
35.70  
357.0  
614  
0.012  
0.002  
00.03  
-000.1  
+3.534  
-34.67  
+101.0  
-594  
10 mV, 50 Hz  
10 mV, 400 Hz  
3.5 V, 50 Hz  
3.5 V, 400 Hz  
35 V, 50 Hz  
35 V, 400 Hz  
350 V, 80 Hz  
600 V, 400 Hz  
0 V  
6
7
8
91  
Right  
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
21  
22  
23  
24  
25  
26  
27  
28  
292  
302  
31  
32  
33  
34  
35  
36  
37  
38  
39  
404  
414  
Center  
Center  
Center  
Center  
Center  
Center  
Center  
Center  
Center  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
Right  
0 V  
+20 mV dc  
-200 mV dc  
+3.5 V  
-35 V dc  
+100 V dc  
-600 V dc  
0.0 µA, dc  
10.0 µA, dc  
-10.0 µA, dc  
200.0 µA, dc  
-200.0 µA, dc  
40.0 µA, 50 Hz  
40.0 µA, 400 Hz  
200.0 µA, 50 Hz  
200.0 µA, 400 Hz  
+2 V dc2  
200 µA dc  
-0.2  
9.7  
-10.3  
197.8  
-202.2  
38.9  
0.2  
10.3  
-9.7  
202.2  
-197.8  
41.1  
200 µA dc  
200 µA dc  
200 µA dc  
200 µA dc  
200 µA ac  
200 µA ac  
38.9  
41.1  
200 µA ac  
195.7  
195.7  
1.980  
3.3973  
-0.5073  
0.0  
204.3  
204.3  
2.020  
3.4633  
-0.4933  
0.2  
200 µA ac  
Continuity  
Continuity, Automatic Selection  
Continuity, Automatic Selection  
400.0 Ω  
+3.43 V dc2  
-0.5 V dc2  
0.0 Ω  
400.0 Ω  
1.0 Ω  
0.8  
1.2  
400.0 Ω  
350.0 Ω  
346.6  
0.990  
34.67  
99.0  
2.873  
34.44  
1.9403  
-0.001 µF  
0.930 µF  
353.4  
1.010  
35.33  
101.0  
2.927  
35.56  
1.9803  
0.001 µF  
0.970 µF  
4.000 kΩ  
1.0 kΩ  
40.00 kΩ  
35 kΩ  
400.0 kΩ  
100 kΩ  
4.000 MΩ  
2.9 MΩ  
40.00 MΩ  
35 MΩ  
, Automatic Selection  
1.000 µF  
1.96 V dc  
0.0 µF  
1.000 µF  
0.95 µF  
1. In the Automatic Selection mode, the Meter uses a low-impedance thermistor (~2.5k for circuit protection and load testing  
(referred to as low-Z input circuitry). When using the 5100B, 5500A, or 5700A to drive the Meter with high voltages, avoid an  
overload/current limit condition by gradually stepping the voltage up (waiting two seconds between each step) from 90.0 V,  
120.0 V, 180.0 V, and 350.0 V at 80 Hz each step.  
2. Calibrator 50divider override.  
3. The dc volts annunciator must be on.  
4. Conducting performance tests of the 400 , 4 k, 40 k, and 1 µF ranges (tests no. 22, 23, 24, 25, 26, 31, and 32) verifies  
that the discrete and integrated circuitry needed to support the other capacitance ranges are working within specifications.  
Therefore, the tests indirectly verify that the Meter will meet specification in the 10 µF, 100 µF, 1000 µF, and 10,000 µF  
ranges.  
Table 4. Temperature Performance Tests  
Input Voltage  
Display Readings  
°C  
°F  
0.0  
0.919 mV  
4.095 mV  
0.0 ± 0.8  
23.0 ± 1.0  
100.0 ± 1.8  
32.0 ± 1.5  
73.4 ± 2.2  
212.0 ± 3.6  
8
 
Calibration Adjustments  
Temperature  
AK-80  
+
5700A  
Red  
Yellow  
Copper  
+
Banana  
Plug  
0
Ice Bath  
ig6f.eps  
Figure 3. Performance Test for Temperature  
Clip to Base Contacts  
+
+
Input Receptacles  
_
+
9V  
Adjust R4 for 3.997-4.003V DC  
BATTERY  
ig1f.eps  
Figure 4. Calibration Adjustment for DC Voltage  
9
 
16  
Multimeter with Temperature  
4. Connect the other end of the thermocouple wire to the  
Meter using a Fluke 80AK Adapter Plug (see Figure 6).  
Allow the instrument reading to settle.  
Temperature  
Note  
Adjust DC voltage (R4) before adjusting  
temperature (R38).  
5. Adjust R38 so that the Meter display has the same  
reading of the ice bath temperature.  
The following describes three temperature calibration  
procedures:  
Lag Bath Calibration  
Calibrating at room temperature. Use this procedure if  
it is not necessary to optimize the Meter’s performance.  
The following adjustment optimizes the Meter’s  
performance at room temperature using a specific  
thermocouple.  
Calibrating in an ice bath. Use this procedure to  
optimize the Meter’s performance at around 0.0°C.  
Calibrating in a lag bath. Use this procedure to  
optimize the Meter’s performance at room temperature.  
1. Put the Meter in the temperature function and °C  
mode.  
2. Insert a thermocouple in a lag bath (see Figure 6).  
Room Temperature Calibration  
The following procedure allows you to make the calibration  
adjustment at room temperature.  
3. Insert a mercury thermometer to the same depth as the  
thermocouple wire to verify that the lag bath  
temperature is at room temperature  
4. Connect the other end of the thermocouple wire to the  
Meter using an 80AK Adapter Plug(see Figure 6).  
Allow the instrument reading to settle.  
1. Allow the Meter to stabilize at room temperature. After  
the temperature stabilizes, do not touch the COM jack.  
2. Turn to temperature function and °C mode.  
5. Adjust R38 so that the Meter display has the same  
reading of the lag bath thermometer.  
3. Observing the correct polarity, connect a copper short  
across the input jacks of the Meter (see Figure 5).  
Replacement Parts  
Replacement parts are listed in Table 4 and shown in  
Figure 1.  
4. Using a temperature probe with the appropriate Meter  
(Fluke 45 or equivalent), measure the temperature of  
the COM input jack (see Figure 5). Be sure to place the  
probe on the COM closest to the pca.  
Table 4. Replacement Parts  
5. Adjust R38 so the Meter display is the same as the  
temperature probe reading.  
Parts  
Part No.  
Window decal  
Case, top  
LCD  
617974  
617966  
643541  
867242  
867247  
618022  
203445  
879031  
696534  
618097  
832246  
885884  
Ice Bath Calibration  
The following procedure optimizes the Meter’s performance  
near 0°C (32°F) using a specific thermocouple.  
Connection elastomeric  
Connection Elastomeric LCD to PCA  
Switch assembly  
1. Put the Meter in the temperature function and °C  
mode.  
Actuator, Switch  
2. Insert a thermocouple in an ice bath (see  
Figure 6).  
Support, Switch Assembly  
Battery  
Case, Bottom assembly  
Screw, THD Form, PH.P.STL, 5014 x .750  
Foot  
3. Insert a mercury thermometer to the same depth as the  
thermocouple wire to verify that the ice bath  
temperature is 0.0°C .  
10  
 
Replacement Parts  
DUAL DISPLAY MULTIMETER  
45  
V
10A  
FUSED  
1000V  
750V  
MAX  
m V  
CAL  
ENABLE  
!
100  
mA  
COM  
dB  
2ND  
V
V
A
A
REL  
500mA  
F
250V  
REF#  
REF  
LOCAL  
LL INPTS  
1kV MX  
AUTO  
COMP  
FREQ  
HOLD  
MN MX  
RATE  
HI  
LO  
THRESH ADDR  
BAUD  
POWER  
Clip to Base Contacts  
8
+
+
_
+
9V  
BATTERY  
R38  
ig4f.eps  
Figure 5. Calibration Adjustment for Temperature Using the 80T-150U  
Thermometer  
Ice or Lag Bath  
Clip to Base Contacts  
+
+
80AK  
+
_
+
R38  
9V  
BATTERY  
Thermocouple  
ig5f.eps  
Figure 6. Calibration Adjustment for Temperature Using an Ice or Lag Bath  
11  
 
16  
Multimeter with Temperature  
12  
 

Everpure Water System EV9333 40 User Manual
Excalibur electronic Games VR15 User Manual
Extron electronic Projector SYSTEM 8 10 PLUS User Manual
Friedrich Air Conditioner CP05A10 User Manual
Frigidaire Air Conditioner FRP9OPTT3R User Manual
Frymaster Fryer SR52 SR62 User Manual
Fujitsu Mouse FKB8735 Series User Manual
GE Convection Oven PK956BM User Manual
GE Microwave Oven GE JVM1850 User Manual
Gigabyte Computer Hardware GO RAMDISK BOX RH User Manual